ISSN:
0935-6304
Keywords:
Open tubular gas chromatography
;
Dynamic headspace analysis
;
Purge and trap
;
Thermal desorption
;
Reinjection
;
Fourier transform infrared spectroscopy
;
Volatile constituents in polymers
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
An analytical system for the analysis of volatiles entrained in polymers is described. This system is based on a thermal desorption oven connected to a cold trap. After enrichment of headspace vapor, trapped material is reinjected and analyzed by open tubular gas chromatography/Fourier transform infrared spectroscopy, OTGC/FTIR. The thermal desorption oven is designed to provide different modes of sample introduction: use of a pyroprobe; insertion of a piece of quartz tubing with applied sample; or syringe injection. Headspace enrichment is carried out in a piece of fused silica capillary tubing filled with glass beads. The trap may be cooled either electrically using Peltier elements or with liquid nitrogen. A six-port rotary valve is used for flow switching between enrichment and reinjection modes. All system parameters, temperatures, and timed events, are controlled from the gas chromatograph. Dynamic headspace analysis is demonstrated as a method for polymer characterization.
Additional Material:
10 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/jhrc.1240110305
Permalink