Anaesthesia
56 (2001), S. 0
ISSN:
1365-2044
Source:
Blackwell Publishing Journal Backfiles 1879-2005
Topics:
Medicine
Notes:
Point-of-care testing of coagulation parameters provides a more rapid assessment of test results compared with laboratory testing. A new coagulation monitor (GEM® PCL, Instrumentation Laboratory, Kirchheim, Germany) was evaluated. Point-of-care data for activated partial thromboplastin time and prothrombin time (expressed as the international normalised ratio) and turn-around-time were compared. Coagulation parameters were compared in the blood of 57 patients with and without heparin therapy. The point-of-care and laboratory test results showed a bias (SD) of − 0.26 (4.55) s for activated partial thromboplastin time and − 0.011 (0.150) s for prothrombin time. The average turn-around-time was 3 min for point-of-care testing vs. 52 min for laboratory testing. We conclude that the reliability of point-of-care testing is sufficient for clinical use.