    Topics: Medicine
    Description / Table of Contents: Summary Trasient asynchronous pacing due to abnormal sensing function is reported in two patients with inserted demand pacemakers during the early phase of acute myocardial infarction. The hazards of the pacemaker induced parasystole with R on T phenomenon in conditions of enhanced electrical instability could be successfully overcome applying overdrive suppression of the inserted pacing system by external chest wall stimulation
    Notes: Zusammenfassung Bei 2 Patienten mit implantiertem Demand-Schrittmacher wurde in der Frühphase eines Myokardinfarktes vorübergehend asynchrone Schrittmachertätigkeit infolge gestörter Detektion beobachtet. Die Gefährdung durch schrittmacherinduzierte Parasystolie mit R auf T Phänomen unter Bedingungen erhöhter elektrischer Instabilität konnte erfolgreich unter Anwendung externer Overdrive-Stimulation beseitigt werden.