ISSN:
1432-2013
Keywords:
Coronary Sinus Blood Flow
;
Thermodilution Technique
;
Mixing of Indicator and Blood
Source:
Springer Online Journal Archives 1860-2000
Topics:
Medicine
Notes:
Summary In hydraulic non-occlusive models, mixing effects were studied with slug-injection and constant-infusion techniques. Measurements were accomplished by means of catheters specially constructed with thermistors. Constant-rate infusion of 5 ml/min of saline at room temperature showed homogeneity of mixing. Good correlation (r=0.936) was found between metered flow ranging from 50–450 ml/min and reciprocal temperature changes. In anesthetized open chest dogs reciprocal temperature changes during constant-infusion thermodilution was compared with coronary sinus flow metered by graduated cylinder and stop-watch. For this purpose all the coronary sinus blood was directed through a tubing system with a Y-connector into the jugular vein. In the range of 50–450 ml/min the metered coronary sinus blood flow and the reciprocal of the observed changes in temperature showed a linear correlation (r=0.949). The reproducibility of the thermodilution method is fairly good. Therefore it is believed that this technique will allow accurate measurement of coronary sinus blood flow.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF00588616