Abstract An MIP-AES element specific detector has been directly coupled to an HPLC separation system by means of hydraulic high-pressure nebulization (HHPN) with an appropriate desolvation unit. This system was applied to Cr(III)/Cr(VI) speciation analysis. The linear dynamic range of Cr(III) and Cr(VI) determination was two orders of magnitude, absolute detection limits were 13 ng for Cr(III) and 18 ng for Cr (VI), respectively. The matrix effect caused by the easily ionizable elements (Na, Ca, K) was significant in real samples.