The electronics for the microchannel plate photomultiplier tubes used in the LIDAR (Light Detection and Ranging) Thomson scattering diagnostic on the JET tokamak are described. It is shown how the microchannel plate photomultiplier tubes are gated with a very small coupling between gate signal and output, and how ≈300 V transient pulses with rise times of 230 ps are clipped to safe levels of ≈40 V.