Library

feed icon rss

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
Filter
  • 1995-1999  (1)
  • 1990-1994  (1)
  • 1995  (1)
  • 1990  (1)
  • 1
    Electronic Resource
    Electronic Resource
    [S.l.] : American Institute of Physics (AIP)
    Journal of Applied Physics 67 (1990), S. 4456-4458 
    ISSN: 1089-7550
    Source: AIP Digital Archive
    Topics: Physics
    Notes: The spin dependence of L2,3 absorption in 5d atoms oriented in a ferromagnetic matrix contains information on the spin density of the empty d-projected states of the absorbing atom. Spin-dependent absorption spectroscopy using circularly polarized synchrotron radiation was applied to study the polarization of the Pt atoms in the binary alloy Co50Pt50 and Pt/Co layered structures, which are promising candidates for magneto-optical recording. The spin-dependent absorption signals for vapor-deposited 250(4 A(ring) Co+18 A(ring) Pt) and 250(6 A(ring) Co+18 A(ring) Pt) multilayers indicate a ferromagnetic coupling on Pt and Co atoms with a significant Pt polarization. This is reduced on average by about 60% with respect to the Pt polarization in the Co50Pt50 alloy. The experimental results are discussed on the basis of spin-polarized band-structure calculations.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
  • 2
    ISSN: 1432-055X
    Keywords: Schlüsselwörter Isofluranpräparate ; Verdampfergenauigkeit ; Vapor 19.n Dräger ; Key words Isoflurane preparations ; Precision of vaporization ; Vaporizer Dräger type 19.n
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Description / Table of Contents: Summary Objectives. The purpose of this study was to check the precision of the Dräger vaporizer model 19.3 when filled with three different preparations of isoflurane. Methods. Six Dräger vaporizers model 19.3 calibrated with forene were filled with forene (Abbott), isoflurane (Lilly) and isoflurane (Pharmacia); gas output was measured by infrared absorption (Irina, Dräger) at vaporizer settings of 0.2, 0.4, 0.6, 0.8, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0 and 5.0 vol%, starting with a fresh gas flow of 2 l/min followed by 4, 6 and 12 l/min. Thus each of the three isoflurane preparations was checked in six different vaporizers and with four different fresh gas flows. Results. Within the concentration range used in clinical practice there was no significant difference in the delivery of the three isoflurane preparations. Each of the six vaporizers produced a controllable and predictable concentration of the three preparations. Conclusion. Vaporizers of Dräger type 19.n calibrated with forene deliver the same predictable concentration of the volatile anaesthetic when filled with isoflurane from Lilly or isoflurane from Pharmacia instead of forene and may be used without impairment in patient safety. In addition, no specific calibration with one of the new isoflurane preparations is required.
    Notes: Zusammenfassung Untersucht wurde die Genauigkeit von mit Forene kalibrierten Isofluranverdampfern der Baureihe 19.3 der Fa. Dräger bei Füllung mit den neuen Präparaten Isofluran-Lilly und Isofluran-Pharmacia; als Vergleichssubstanz diente Forene der Fa. Abbott. Die Prüfung erfolgte an 6 Vaporen bei Konzentrationseinstellungen von 0,2, 0,4, 0,6, 0,8, 1,0, 1,5, 2,0, 2,5, 3,0, 3,5, 4,0 und 5,0 Vol.-%; zuerst bei einem Fluß von 2 l Luft/min, dann in weiteren Meßreihen bei 4, 6 und 12 l/min. Die Isoflurankonzentrationen wurden mit Infrarotabsorptionstechnik gemessen (Irina, Dräger). Insgesamt wurde jede Prüfsubstanz in 6 verschiedenen Vaporen bei 4 unterschiedlichen Einstellungen des Frischgasflusses untersucht. Bei der Verdampfung der 3 Isofluranpräparate unterschiedlicher Hersteller ergab sich im klinisch üblichen Dosierungsbereich kein signifikanter Unterschied. Diese Aussage einer für alle Prüfsubstanzen gleichen Narkosemittelverdampfung läßt sich auf alle Dräger-Vaporen der Bauart 19.n übertragen, da die Konstruktionsunterschiede lediglich das Dräger-Stecksystem „Interlock“, einen Dichtungsschutz und die Überlaufsicherung betreffen. Aus der vorliegenden Untersuchung ergibt sich, daß mit Forene kalibrierte Dräger-Vapore der Baureihe 19.n die neuen Präparate Isofluran-Lilly und Isofluran-Pharmacia mit gleicher Konzentrationsgenauigkeit abgeben und daher ohne Gefahr für den Patienten hiermit gefüllt werden dürfen. Eine Neukalibrierung der Verdampfer mit diesen Präparaten ist nicht erforderlich.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...