ISSN:
0269-3879
Schlagwort(e):
Chemistry
;
Analytical Chemistry and Spectroscopy
Quelle:
Wiley InterScience Backfile Collection 1832-2000
Thema:
Medizin
Notizen:
An isocratic reversed-phase high performance liquid chromatographic method has been developed for the determination of dobutamine in the plasma of dialysed patients. A solid phase extraction method with a Sep-Pak C18 cartridge was used to isolate the drug and isoxsuprine (internal standard) from plasma. The separation was carried out on an ODS-Hypersil column with 0.1 M phosphate buffer: acetonitrile: methanol (72 : 20 : 8 v/v/v) as the mobile phase. The recovery of dobutamine added to plasma by the extraction procedure was 87 ± 2.3% (mean ± SD). The accuracy and reproducibility of the method were within acceptable limits over the concentration range 0-1000 ng/ mL. Quantification was by fluorescence detection at 275 nm excitation and 310 nm emission wavelengths with a detection limit of 5 ng/ mL for dobutamine. This procedure was applied to ascertain the pharamacokinetics of dobutamine infusion in nine patients with cardiogenic shock and end-stage renal disease undergoing haemodialysis.
Zusätzliches Material:
3 Ill.
Materialart:
Digitale Medien
URL:
http://dx.doi.org/10.1002/bmc.1130080613