Library

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    Woodbury, NY : American Institute of Physics (AIP)
    Applied Physics Letters 54 (1989), S. 2651-2653 
    ISSN: 1077-3118
    Source: AIP Digital Archive
    Topics: Physics
    Notes: A new atomic force microscope, which combines a microfabricated cantilever with an optical lever detection system, now makes it possible to measure the absolute force applied by a tip on a surface. This absolute force has been measured as a function of distance (=position of the surface) in air and water over a range of 600 nm. In the absolute force versus distance curves there are two transitions from touching the surface to a total release in air caused by van der Waals interaction and surface tension. One transition is due to lifting off the surface; the other is due to lifting out of an adsorbed layer on the surface. In water there is just one transition due to lifting off the surface. There is also a transition in air and water when the totally released tip is pulled down to touch the surface as the surface and tip are brought together. Based on the force versus distance curves, we propose a procedure to set the lowest possible imaging force. It can now be as low as 10−9 N or less in water and 10−7 N in air.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗