Library

feed icon rss

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    ISSN: 1432-055X
    Keywords: Schlüsselwörter Heat and Moisture Exchanger ; HME ; aktive Befeuchter ; Atemgasklimatisierung ; Wärmetransport ; Wärmeverluste ; Key words Heat and moisture exchanger (HME) ; Air conditioning ; Heated humidifiers (HH) ; Heat transport ; Pulmonary heat loss ; Body rewarming
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Description / Table of Contents: Abstract Heated humidifiers (HH) as well as heat and moisture exchangers (HME) are commonly used in intubated patients as air-conditioning devices to raise the moisture content of the air, thus preventing mucosal damage and heat loss resulting from ventilation with dry inspired gases. In contrary to HME, HH are able to add heat and moisture to the inspired air in surplus, which is often stressed as an advantage in warming hypothermic patients or reducing major heat losses, e.g., during long operations. The impact of air conditioning on the energy balance of man was calculated comparing HME and HH. Methods. The efficiency of a HME (Medisize Hygrovent) and a HH (Fisher & Paykel MR 730) was evaluated in a mechanically ventilated lung model simulating the physiological heat and humidity conditions of the upper airways. The gas flow from the central supply was dry; the model temperature varied between 32 and 40 °C. By using a HH in the inspiratory limb, a circle system was simulated with water-saturated inspired air at room temperature. The water content of the ventilated air was determined at the tracheal tube connection using a fast, high-resolution humidity meter and was compared with the moisture return of the HME. The energy balance was calculated according to thermodynamic laws. Results. Both HME and HH were able to create physiological heat and humidity conditions in the airways. With the normothermic patient model, the moisture return of the HME was equal to that of the HH set at 34 °C. Increasing the heating temperature resulted only in reduced water loss from the lung; heat and water input in the normothermic model was not possible. This was only effective with almost negligible amounts under hypothermic patient model conditions. Discussion. The water content in the inspired and expired air is the most important parameter for estimating pulmonary heat loss in mechanically ventilated patients. In adults (minute volume ∼7 1/min) the main fraction of pulmonary heat loss results from water evaporation from the airways (∼6 kcal/h), whereas the heat loss due to convection is negligible (∼1.2 kcal/h). In intubated patients ventilated with dry air, the heat loss increases to ∼8 kcal/h due to greater water evaporation from the airways. Both HME and HH are able to reduce the pulmonary heat loss to 1–2 kcal/h. In normothermic as well as hypothermic patients, HH do not offer significant advantages in heat balance compared to effective HME. In conclusion, air conditioning in intubated patients is neither a powerful too for maintaining body temperature during long-lasting anaesthesia nor a sufficient method of warming hypothermic patients in intensive care units.
    Notes: Zusammenfassung Pulmonal bedingte Wärmeverluste bei Beatmung mit trockenen und kalten Atemgasen sind in der perioperativen Phase ebenso wie in der Intensivmedizin durch atemgasklimatisierende Maßnahmen weitgehend vermeidbar. Im Vergleich zu Energieverlusten durch Radiation, Konvektion und Evaporation von Wärme und Wasser von der Körperoberfläche und aus geöffneten Körperhöhlen ist die Wärmetransportkapazität der Atemluft jedoch gering. Die Kompensation hoher perioperativer Wärmeverluste sowie die Wiedererwärmung hypothermer Patienten ist auch durch Beatmung mit überkörperwarmen Inspirationsgasen nicht möglich. Aktive Befeuchtungssysteme (heated humidifier, HH) bieten somit aus energetischer Sicht keine wesentlichen Vorteile gegenüber leistungsfähigen Wärme- und Feuchtigkeitstauschern (heat and moisture exchanger, HME).
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
  • 2
    ISSN: 1432-055X
    Keywords: Schlüsselwörter Hypothermie ; Wärmeverfahren ; Hämofiltration ; Key words Hypothermia ; Rewarming techniques ; Haemofiltration
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Description / Table of Contents: Abstract We report the case of a 35-year-old male who was admitted to the intensive care unit because of somnolence due to accidental hypothermia. Initial examination showed a Glasgow coma score of 10 and a rectal temperature of 27.4° C. Because of stable circulatory conditions, there was no mandatory indication for rewarming by means of cardiopulmonary bypass. We rewarmed the patient with an extracorporeal veno-venous haemofiltration device combined with a countercurrent fluid warmer. An average increase in body temperature of 1.34° Ch−1 could be obtained. We conclude that the described technique represents an effective and well-controllable method for treatment of hypothermia in patients with stable haemodynamic conditions. Because of the availability of the required equipment, this method can also be practised in hospitals without cardiac surgical departments and cardiopulmonary bypass facilities.
    Notes: Zusammenfassung Beschrieben wird die Kasuistik eines 35jährigen Patienten, der wegen akzidenteller Hypothermie intensivmedizinisch behandelt wurde. Aufgrund unklarer Somnolenz wurde der Patient vom ärztlichen Notdienst zur stationären Behandlung eingewiesen. Bei Aufnahme zeigte sich eine deutliche Bewußtseinsstörung sowie eine Hypothermie mit einer Rektaltemperatur von 27,4° C. Der Patient wurde zur Wiedererwärmung auf die Intensivstation verbracht. Die Reduktion der hämodynamischen Parameter entsprach dem Grad der Hypothermie, die stabile kardiozirkulatorische Situation ergab keine zwingende Indikation zur Wiedererwärmung mit der Herz-Lungen-Maschine. Statt dessen wurde die aktive Anhebung der Körpertemperatur durch einen extrakorporalen, veno-venösen Wärmekreislauf durchgeführt, der aus einer Hämofiltrationspumpe und einem Wärmetauscher bestand. Mit dieser Methode wurde ein durchschnittlicher Temperaturanstieg von 1,34° C/h erreicht. Mit dem beschriebenen Verfahren wird eine effektive und gut steuerbare Methode zur Wiedererwärmung hypothermer Patienten vorgestellt, die – im Gegensatz zum kardiopulmonalen Bypass – aufgrund des geringen apparativen Aufwands auch an Kliniken mit niedriger Versorgungsstufe durchgeführt werden kann.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
  • 3
    ISSN: 1432-055X
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Notes: Zusammenfassung. Die Befeuchtungsleistung unterschiedlicher Wärme- und Feuchtigkeitstauscher (DAR Hygrobac, Gibeck HumidVent 2P, Pall BB 22-15 T und BB 100) zur Klimatisierung der Atemgase wurde im beatmeten Lungenmodell unter standardisierten Bedingungen sowie bei 25 intubierten Patienten untersucht. Mit Hilfe eines neuartigen, hochauflösenden Feuchtesensors konnte nachgewiesen werden, daß die Klimatisierungsleistung von Wärme- und Feuchtigkeitstauschern (HME, Heat and Moisture Exchanger) unter Nichtrückatmungsbedingungen nicht nur durch Messung der inspiratorischen Feuchte direkt am Tubus bestimmt werden kann, sondern einfacher und genauer durch Messung des exspiratorischen Wassergehalts in der Abluft des Respirators. Diese Meßmethodik ist nicht-invasiv und erlaubt jederzeit die Bestimmung der Befeuchtungsleistung des verwendeten HME in der spezifischen Beatmungssituation. Anders als bei der schwierigen Interpretation der tubusnah ermittelten Feuchteparameter ist durch die Bestimmung des Wassergehalts der Exspirationsluft am Respiratorauslaß eine eindeutige Beurteilung der Befeuchtungsleistung des HME durch Vergleich mit den physiologischen Wärme- und Feuchtigkeitsverhältnissen in den Atemwegen möglich. Durch Beschichtung des Innenmaterials mit hygroskopischen Substanzen kann die Wasserbindungskapazität von HME soweit erhöht werden, daß– gemessen an physiologischen Gegebenheiten – adäquate Befeuchtungsleistungen erreicht werden können. HME stellen damit prinzipiell eine gute Alternative zu aktiven Befeuchtungssystemen dar.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
  • 4
    ISSN: 1432-055X
    Keywords: Schlüsselwörter Infusionswärmung ; Durchflußwärmer ; Hypothermie ; Wärmeverluste ; Key words Body temperature ; Hypothermia ; Infusion ; Blood transfusion ; Equipment design
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Description / Table of Contents: Abstract Heat loses during surgery occur mainly to the environment and due to infusions and irrigations. Infusions given at room temperature account for a great deal of the total heat deficit during major operations, e.g., the infusion of 53 ml/kg 20° C fluid leads to a loss of 1° C in mean body temperature. Hence, heating i.v. fluids will add to the effect of other measures aimed at reducing heat loss to the environment. We investigated the efficacy of different warming methods for i.v. fluids in an experimental model by measuring the temperature at the end of the delivery line. Methods. The following in-line warmers were studied: Hotline HL-90 and System H-250/heat exchanger D-50 (Level 1 Technologies, Marshfield, USA), Astotherm IFT 260 (Stihler Elektronic GmbH, Stuttgart, Germany), RSLB 30 H Gamida (Productions Hospitalieres Francaises, Eaubonne, France), Bair Hugger 241/Modell 500 Prototype (Augustine Medical, Eden Prairie, USA). They were compared with prewarming infusions (39° C) only using the Clinitherm S (Labor Technik Barkey GmbH, Bielefeld, Germany) and prewarming with “active insulation” of the delivery line using the Autotherm/Autoline system (Labor Technik Barkey GmbH, Bielefeld, Germany). We investigated the influence of four variables on the efficacy of warming: (1) flow rate (50–15,000 ml/h); (2) ambient temperature (20° C and 25° C); (3) infusion bag temperature (6° C, 20° C, and 39° C); and (4) length of infusion system downstream from the heat exchanger. Fluid temperatures were measured using thermistors of 1 mm diameter (Modell YSI 520, Yellow Springs Instruments Co., Yellow Springs, USA) incorporated into 3-way stopcocks. Temperatures were recorded using Hellige temperature monitors (Hellige GmbH, Freiburg im Breisgau, Germany) and the signals were collected at 10 Hz through an AD converter and averaged over 1 min. Flows were calculated by timed collection into calibrated cylinders; 10 to 12 different flow rates were taken to define one temperature/flow plot. Effective warming was defined as a temperature 〉33° C at the end of the infusion line. Results. At high flow rates (〉2,500 ml/h) using 20° C fluids at 20° C ambient temperature, the H-250/D-50 system gave the highest temperatures throughout the range and showed effective warming from 1,300 ml/h on over the entire range tested (35° C at 17,000 ml/h) compared to the RSLB 30 H Gamida system (3,000–18,000 ml/h) (Fig. 2). This difference in performance was almost abolished with fluids at 6° C (Fig. 4). Similar efficacy could be reached by using prewarmed infusions that gave effective warming at 〉2,000 ml/h and reached 39° C at 13,000 ml/h. Prewarmed infusions could be used effectively down to 〉80 ml/h applying “active insulation” (Autotherm/Autoline) to the whole infusion system. The Hotline HL-90 (50–4,700 ml/h) appeared to be the most effective in-line warmer in the low (〈250 ml/h) and middle (250–2,500 ml/h) flow range, followed by the Astotherm IFT 260 (400–4,000 ml/h), but only if used with a length of 40 cm down-stream from the heat exchanger (Fig. 1). Increasing this distance to 145 cm markedly reduced its efficacy below the range of 2,000 ml/min (1,200–3,000 ml/h) (Fig. 5). The Bair Hugger 241 Prototype showed a narrow effective range (700–1,300 ml/h) that could be extended beyond 1,300 ml/h by the use of prewarmed infusions (Figs. 1 and 3). The performance for 6° C solutions and ambient temperatures of 25° C are given in Fig. 4 and Table 1. Conclusions. The importance of infusion warming increases with the amount of fluid given. In general, the infusion bag temperature only influenced the efficacy of in-line warmers within the high-flow range, challenging the performance of the heat exchanger. The length of uninsulated i.v. line downstream from the heat exchanger influenced the efficacy within the low- and middle-flow range, as did the room temperature. Prewarmed solutions can be infused very effectively within the high-flow range. This efficiency can be preserved down to the low-flow range by using “active insulation” of the infusion system. In-line warming is essential for emergency and rapid massive transfusions.
    Notes: Zusammenfassung Eine Hypothermie gehört zu den häufigsten Komplikationen in der perioperativen Phase. Eine ihrer Ursachen liegt in der Applikation unzureichend erwärmter Blut- und Infusionslösungen. Es wurde die Effektivität verschiedener Erwärmungsverfahren untersucht: 1) Vorwärmung von Infusionen (39° C) und 2) Verschiedene Durchflußwärmer (System H-250 ® /D-50 ® , RSLB 30 H Gamida ® , Hotline ® HL-90, Autotherm ® /Autoline ® , Astotherm ® IFT 260, Bair Hugger ® 241 Prototyp). Als Effektivitätsgrenze wurde eine patientennahe Infusionstemperatur von ≥33° C definiert. Variiert wurden a) Flußrate (50–15000 ml/h), b) Ausgangstemperatur der Infusion (6, 20 und 39° C), c) Länge des Infusionssystems nach dem Wärmetauscher und d) Umgebungstemperatur (20 und 25° C). Für eine Raum- und Infusionstemperatur von 20° C wurden folgende effektive Arbeitsbereiche gefunden: System H-250 ® /D-50 ® 1300- bis mindestens 17000 ml/min, RSLB 30 H Gamida ® 3000–18000 ml/min, Hotline ® HL-90 50–4700 ml/min, Astotherm ® IFT 260 Infusionssystemlänge 40 cm: 400–4000 ml/min, Infusionssystemlänge 145 cm: 1200–3000 ml/min, Bair Hugger ® 241 700–1300 ml/min, Autotherm ® /Autoline ® mit vorgewärmten Infusionen (39° C) 〉80 ml/min, vorgewärmte Infusionen (39° C) ohne „aktive Isolation“ 〉2000 ml/min. Für die getesteten Variablen gilt: Eine geringe Ausgangstemperatur der Infusion reduziert nur im hohen Flußbereich die Effektivität der Wärmer. Je niedriger Flußrate und Umgebungstemperatur sind und je länger das Infusionssystem nach dem Wärmetauscher ist, desto größer wird der Temperaturverlust auf dem Weg zum Patienten. Bis zu einem Infusionsfluß von 2000 ml/h ist eine effektive Infusionswärmung alleine durch Vorwärmung (39° C) generell nicht möglich.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
  • 5
    ISSN: 1432-055X
    Keywords: Schlüsselwörter Atemgasklimatisierung ; Aktive Befeuchtungssysteme ; HME ; Heat and Moisture Exchanger ; Künstliche Nasen ; Wärme- und Feuchtigkeitstauscher ; Key words Air-conditioning ; Artificial nose ; Heat and moisture exchanger ; Heated humidifier ; HME
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Description / Table of Contents: Abstract Heat and moisture exchangers (HME) are used as artificial noses for intubated patients to prevent damage resulting from dry and cold inspired gases. HME collect a large fraction of the heat and moisture of the expired air, adding them to the subsequent inspired breath. In a prospective clinical study the air-conditioning capacity of a heated humidifier was compared with a hygroscopic HME. Methods. The water content of the ventilated air of 49 intensive care patients requiring artificial ventilation with tidal volumes between 440 and 1,190 ml (mean 658±148 ml) was examined. Each patient was ventilated in sequence with an HME (DAR Hygrobac S) and a heated humidifier (Fisher & Paykel MR 630 B). The temperature of the air in the inspiratory limb was maintained at 34 °C. The water content of the ventilated air was determined under steady-state conditions directly at the tracheal tube or between tracheal tube and HME using a new, high-resolution humidity meter. The results were compared with the absolute water loss of the exhaled air at the gas outlet of the ventilator as an expression of the water loss from the lower airways. Airway resistance was calculated by a standard formula. The daily running costs for both HME and heated humidifier were estimated. Results and discussion. Moisture retention was equivalent in both the HME and the heated humidifier (33.7±1.85 bzw. 34.1±2.62 mgH2O/l). These data show that modern HMEs are able to maintain physiological air-conditioning even in long-term ventilated patients. The small increase in airway resistance associated with HMEs (3.1±2.5 mbar/l·s) has to be noted in difficult weaning procedures. Both labour and costs per day are significantly less with HMEs (8.60 vs. 21.70 DM).
    Notes: Zusammenfassung In einer prospektiven klinischen Studie an 49 intubierten und kontrolliert beatmeten Patienten wurde gezeigt, daß die inspiratorischen Befeuchtungsleistungen des HME DAR Hygrobac S und des aktiven Befeuchters Fisher & Paykel MR 630 B äquivalent sind (33,7± 1,85 vs. 34,1±2,62 mgH 2 O/l). Bei bestimmungsgemäßem Einsatz (tubusnahe Plazierung bzw. inspiratorische Atemgastemperatur von 34 °C) und Tidalvolumina zwischen 440 und 1190 ml (Mittel 658±148 ml) gewährleisteten beide Systeme eine physiologische Klimatisierung der Atemluft. Die geringgradige Erhöhung der inspiratorischen Resistance durch die HME (3,1±2,5 mbar/l/s) muß ggf. bei der schwierigen Respiratorentwöhnung mit berücksichtigt werden. Die Verwendung von HME ist erheblich kostengünstiger und weniger personalintensiv als der Einsatz von aktiven Befeuchtungssystemen. Leistungsstarke HME stellen damit eine gute Alternative zu aktiven Befeuchtungssystemen in der Intensivbeatmung dar.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
  • 6
    ISSN: 1432-055X
    Keywords: Schlüsselwörter Fehlerquellen ; Gefahren ; Motorspritzenpumpen ; Infusionssysteme ; Key words Source of faults ; Dangers ; Syringe drivers ; Infusion systems
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Description / Table of Contents: Abstract Syringe drivers are used in anaesthetics, intensive care and emergency medicine to deliver small volumes of highly potent drugs with continuous, constant and reproducible flow. For early recognition of interruptions of the drug delivery caused by occlusion of the infusion system, an alarm is triggered as soon as the system pressure exceeds a certain limit. The sensitivity of this alarm depends on the flow rate, type-specific cut-off pressure and the elastic parameters of the infusion system. The sudden release of pressure built up in the system after occlusion occurred can cause delivery of an uncontrolled drug bolus and hence an additional hazard. Methods: Six syringe drivers that are widely used in clinical practice were tested for alarm delay and bolus delivery in the event of an occlusion in the system. First, the alarm pressures at flow rates of 10, 50 and 100 ml/h were measured. Then the alarm delay time and bolus volumes post-occlusion were assessed, using a basic infusion system (syringe+single infusion set). Finally, several alterations to the system like extension, tap battery with germ filters or branching were made and their impact on alarm delay and bolus volume measured. Results: Because of the great differences in alarm pressures between the devices tested, there were marked differences in the alarm delay at same flow rates. Predictably, there was an indirect proportional link between alarm delay and flow rate. Using the basic infusion system, alarm delays between 23 s and 143 min were measured. In two of the tested syringe drivers, a pressure-release mechanism is activated with the pressure alarm, which prevented bolus application. In the other devices, release of the pressure in the occluded system caused boli of 0.5–7 ml. Variations in the actual syringe volume and insertion of a second connection tube had no impact on alarm delay and bolus volume. Tap batteries, parallel running syringe drivers or trapped air in the system, however, caused marked increase in both alarm delay (107%) and bolus volume (+147%). Discussion: Unidentified occlusions of the system cause grave malfunctioning of syringe drivers. While applying highly potent drugs, the discontinuation of drug delivery with subsequent bolus application can cause vital danger to the patient. As a result of the significant time delays in the pressure alarms, the devices tested do not provide sufficient protection against unrecognized system occlusion. Syringe drivers with adjustable alarm pressure can be set close to the actual infusion pressure. A further important point is that one should aim at a reduction in the elastic properties of the infusion set because of the great impact on alarm delay and bolus volume.
    Notes: Zusammenfassung Im Modell wurde die Gerätefunktion verschiedener gebräuchlicher Motorspritzenpumpen bei akzidentellem Verschluß der Infusionsleitung untersucht. Dabei zeigte sich, daß insbesondere Geräte älterer Bauart ohne die Möglichkeit zur Variation der Druckabschaltgrenzen erhebliche Alarmverzögerungen aufweisen. Bei Verwendung eines einfachen Infusionssystems (50 ml-Spritze mit 200 cm-Druckinfusionsleitung) wurden in Abhängigkeit der eingestellten Förderrate Verzögerungszeiten zwischen 23 s und 143 min gemessen. Die meisten der untersuchten Pumpen gaben nach Druckentlastung Flüssigkeitsboli ab, deren Volumen zwischen 0,5 und 1,7 ml lag. Einzelne Modifikationen am Infusionssystem (zusätzliche parallelgeschaltete Pumpen, Lufteinschlüsse in den Spritzen, zugeschaltete Hahnbänke und Infusionsfilter etc.) hatten zum Teil erhebliche Zunahmen der Verzögerungszeiten (bis zu 107%) und der Bolusvolumina (bis zu 147%) zur Folge. Aus den Ergebnissen wird gefolgert, daß Motorspritzenpumpen zur Applikation hochpotenter kreislaufwirksamer Medikamente bei niedrigen Förderraten über variabel einstellbare Abschaltdruckgrenzen verfügen sollten. Unerwünschte Bolusapplikationen müssen ausgeschlossen sein.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
  • 7
    Electronic Resource
    Electronic Resource
    Springer
    Der Anaesthesist 44 (1995), S. 116-118 
    ISSN: 1432-055X
    Keywords: Schlüsselwörter Swan-Ganz-Katheter ; Komplikationen ; Herzchirurgie ; Röntgendurchleuchtung ; Key words Swan-Ganz catheter ; Complications ; Cardiac surgery ; Fluoroscopy
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Description / Table of Contents: A 58-year-old man, admitted with one-vessel coronary disease and aortic valve insufficiency, was scheduled for cardiac surgery. He underwent single coronary bypass grafting and aorta ascendens replacement by a composite graft. During surgery he was monitored by a Swan-Ganz catheter, which had been inserted uneventfully via the right internal jugular vein. Six hours after completion of surgery we were unable to remove this catheter. Fluoroscopy showed intracardiac entrapment of the catheter. Rethoracotomy was performed, revealing entrapment by a suture. The purse-string suture was released, and the catheter was withdrawn without resistance.
    Notes: Zusammenfassung Bei einem 58jährigen Patienten wurden ein 1fach Koronarbypass und ein Aorta ascendens-Ersatz durch einen klappentragenden Conduit vorgenommen. Der vor dem Eingriff eingeschwemmte Swan-Ganz-Katheter ließ sich postoperativ nicht mehr entfernen. In der Röntgendurchleuchtung fand sich eine Bewegung der Herzsilhouette synchron zum Zug am Katheter. Als Ursache stellte sich eine intrakardiale Fixierung durch eine zu tief gefaßte Naht zur Versorgung der Vent-Einstichstelle heraus, die nach Rethorakotomie gelöst werden mußte. Der Katheter ließ sich danach problemlos entfernen.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
  • 8
    ISSN: 1432-055X
    Keywords: Schlüsselwörter Atemgasklimatisierung ; Beheizte Patientenschläuche ; Koaxiales Patientenschlauchsystem ; Minimal-flow-Anästhesie ; Niedrigflussnarkosen ; Key words Anaesthetic gas climatisation ; Coaxial hose system ; Heated breathing hoses ; Low flow anaesthesia ; Minimal flow anaesthesia
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Description / Table of Contents: Abstract Background. During general anaesthesia gas climate significantly is improved by performance of low flow techniques. Gas climatisation, however, markedly also will be influenced by the temperature loss at, and corresponding water condensation within the hoses, factors which are related to the technical design and material of the patient hose system. The objective of this prospective study was to investigate 1. how anaesthetic gas climatisation during minimal flow anaesthesia is influenced by the technical design of different breathing hose systems in clinical practice. 2. to investigate, whether a sufficient gas climatisation also can be gained with higher fresh gas flows if that hose system is used, proven beforehand to optimally warming and humidifying the anaesthetic gases. Methods. Three different systems, a conventional two-limb hosing consisting of smooth silicone hoses, a coaxial hosing, and a hosing consisting of actively heated breathing hoses, attached to a Dräger Cicero EM anaesthesia machine, were used during minimal flow anaesthesia with a fresh gas flow of 0.5 l/min. Gas temperature and absolute humidity were measured at the tapered connection between the inspiratory limb and the breathing system as well as at its connection to the endotracheal tube. The best gas climatisation was observed if heated breathing hoses were used. Thus, using this hosing, additionally gas temperature and humidity in the inspiratory limb were taken at fresh gas flow rates of 1.0, 2.0 and 4.4 l/min respectively. Measurements were performed in all groups at all general anaesthesias lasting at least 45 minutes during the lists of eight different days each. Results. In minimal flow anaesthesia, with all hose systems likewise, generally an absolute humidity between 17 to 30 mgH2O/l is reached at the endotracheal tube's connector during the course of the list. Only in the first cases of the day there was a short delay of 15 to 30 minutes before reaching a humidity of at least 17 mgH2O/l. Only with heated hoses, however, humidity frequently even exceeded 30 mgH2O/l. If conventional or coaxial hosings were used, during minimal flow anaesthesia gas temperatures in an acceptable range between 23 to 30 °C were measured at the tube connector. With heated hoses, however, warming of the gases was excellent with gas temperatures betwen 28 to 32 °C. In minimal flow anaesthesia climatisation of the anaesthetic gases proved to be best if heated hoses were used. Thus, using heated hose systems another three trials with increasing fresh gas flow rates of 1.0, 2.0 and 4.4 l/min respectively were performed. Whereas climatisation of the anaesthetic gases still was found to be optimal with a fresh gas flow of 1.0 l/min, the humidity dropped drastically to values lower than 17 mgH2O/l at 2.0 l/min and even down to 10 mgH2O/l at a flow rate of 4.4 l/min. Gas temperatures, however, turned out to be independent of the flow and remained at 28–32°C, even at a flow as high as 4.4 l/min. Conclusions. Using conventional hose systems and coaxial hosings acceptable, but not optimal climatisation of the anaesthetic gases can be gained if minimal flow anaesthesia is performed. The use of a coaxial hose system seems to lead to improved climatisation in long lasting procedures only. In routine clinical practice, however, conventional and coaxial hose systems are similar in respect to the climatisation of breathing gases. Heated breathing hoses performed markedly better in terms of climatisation of the breathing gas than the coaxial and the conventional hose system. With this hosing not only sufficient but optimal moisture and temperature values are realized. Optimal climatisation, however, only can be gained if low flow anesthetic techniques with fresh gas flows equal or less than 1 l/min are performed. With higher fresh gas flow rates the humidity decreases markedly while high gas temperatures are maintained. It seems justified to assume, that ventilation with warm but dry gases may result in increasingly drying out of the respiratory epithelium of the lower air ways. Heated hoses only should be used if low flow anaesthetic techniques are performed. While moisture content of the breathing gases mainly is influenced by the fresh gas flow rate, temperature mainly is depending on the convectional loss of heat at the inspiratory limb of the hosing.
    Notes: Zusamenfassung Fragestellung. Die Klimatisierung der Narkosegase wird in erheblichem Maß vom Frischgasfluss beeinflusst und ist bei Durchführung von Niedrigflussnarkosen besser als bei hohem Flow. Feuchte und Wärme der Gase werden darüber hinaus aber auch von den Wärmeverlusten und entsprechender Wasserkondensation in den Atemschläuchen beeinflusst, die ihrerseits vom technischen Design und den Materialeigenschaften der Atemschläuche abhängen. Mit dieser prospektiven Studie sollte untersucht werden, 1. inwiefern bei Durchführung von Niedrigflussnarkosen die Klimatisierung der Narkosegase im klinischen Routinebterieb durch das technische Design der Schlauchsysteme beeinflusst wird. Des weiteren sollte 2. untersucht werden, ob bei Einsatz des Schlauchsystems, mit dem optimale Klimatisierungseffekte erreicht werden, diese sich auch bei Erhöhung der Frischgasflows noch realisieren lassen. Methodik. Drei unterschiedliche Schlauchsysteme, das konventionelle Doppelschlauchsystem, ein koaxiales Schlauchsystem und ein Doppelschlauchsystem mit beheizten Schläuchen wurden bei der Durchführung von Minimal-flow-Anästhesien mit einem Frischgasfluss von 0,5 l/min an einem Dräger Cicero EM Narkosegerät eingesetzt. Die Atemgastemperatur und die absolute Feuchte im Inspirationsschenkel des jeweiligen Schlauchsystems wurden sowohl Atemsystem- als auch tubuskonnektornah gemessen. Die besten Klimatisierungseffekte wurden bei Einsatz der beheizten Schläuche erreicht. Mit diesem Schlauchsystem wurden deshalb weitere Messungen der inspiratorischen Atemgasfeuchte und -wärme mit Frischgasflows von 1,0, 2,0 und 4,4 l/min durchgeführt. Die Messungen in allen Gruppen erfolgten während der Durchführung von Inhalationsnarkosen von zumindest 45 min Dauer an jeweils acht Arbeitstagen. Ergebnisse. Bei der Durchführung von Minimal-flow-Narkosen wurden gleichermaßen während des Ablaufs der Arbeitstage mit allen Schlauchsystemen tubuskonnektornah Feuchtewerte zwischen 17 und 30 mgH2O/l erreicht. Nur während der ersten Narkosen zu Beginn der Arbeitstage wurde eine kurze zeitliche Verzögerung von 15 bis 30 min bis zum Erreichen einer Feuchte von zumindest 17 mgH2O/l beobachtet. Bei Einsatz der beheizten Schlauchsysteme wurden gehäuft auch Feuchtewerte über 30 mgH2O/l gemessen. Die tubuskonnektornah gemessenen Atemgastemperaturen waren bei Einsatz des konventionellen und des koaxialen Schlauchsystems während der Durchführung von Minmal-flow-Narkosen mit Werten zwischen 23 und 30°C in einem akzeptablen Bereich. Bei Einsatz der beheizten Schläuche aber waren die inspiratorischen Atemgastemperaturen mit Werten zwischen 28 bis 32 °C deutlich höher und nahezu optimal. Während der Durchführung von Minimal-flow-Narkosen waren somit bei Einsatz beheizter Schlauchsysteme die beste Klimatisierung der Atemgase zu erreichen. Mit diesem Atemsystem wurden deshalb weitere Messreihen mit zunehmenden Frischgasflows von 1,0, 2,0 und 4,4 l/min durchgeführt. Während bei einem Frischgasfluss von 1,0 l/min die Atemgasklimatisierung weiterhin optimal ist, fällt bei weiterer Steigerung des Frischgasflows auf 2,0 l/min die Feuchte drastisch auf Werte unter 17 mgH2O/l, bei einem Flow von 4,4 l/min sogar auf Werte um nur 10 mgH2O/l ab. Dagegen erwiesen sich die Atemgastemperaturen bei Einsatz beheizter Schläuche als weitestgehend flowunabhängig und lagen immer, auch bei einem Flow von 4,4 l/min, um 28–32°C. Schlussfolgerungen. Mit konventionellen und koaxialen Schlauchsystemen wird bei der Durchführung von Minimal-flow-Anästhesien eine zwar nicht optimale, aber ausreichende Atemgasklimatisierung erreicht. Der Einsatz koaxialer Systeme scheint im Vergleich zum konventionellen Doppelschlauchsystem nur bei langdauernden Narkosen zu einer weiteren Verbesserung der Klimatisierung zu führen. Während des klinischen Routinebetriebs sind beide Systeme bezüglich der Klimatisierungseffekte als gleichwertig anzusehen. Bei Einsatz beheizter Schläuche ist unter den Bedingungen der Minimal-flow-Anästhesie die Klimatisierung der Atemgase deutlich besser als bei Einsatz konventioneller oder koaxialer Schlauchsysteme. Es werden nicht nur ausreichende, sondern gar optimale Feuchte- und Wärmewerte erreicht. Diese optimale Klimatisierung wird aber nur bei Durchführung von Niedrigflussverfahren realisiert, dann also, wenn der Frischgasflow nicht größer als 1 l/min ist. Bei höheren Frischgasflows nimmt die Feuchte der Atemgase drastisch ab, während die Atemgastemperaturen hoch bleiben. Es ist anzunehmen, dass die Beatmung mit warmem und trockenem Atemgas zu verstärkter Austrocknung des Atemwegsepithels der unteren Atemwege führt. Beheizte Schläuche sollten deshalb nur bei Durchführung von Niedrigflussnarkosen eingesetzt werden. Während die Feuchte der Atemgase im wesentlichen von der Frischgasflussrate bestimmt wird, hängt die Atemgastemperatur im wesentlichen vom Wärmeverlust am Inspirationsschenkel des Patientenschlauchsystems ab.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
  • 9
    ISSN: 1432-1238
    Keywords: Key words Accidental hypothermia ; Rewarming strategy ; Critical care ; Extracorporeal circulation
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Notes: Abstract Objective: To study a rewarming strategy for patients with severe accidental hypothermia using a simple veno-venous bypass in combination with a convective air warmer. Setting: Eighteen beds in a university hospital intensive care unit. Patients: Four adults admitted with a core temperature less than 30 °C. Hypothermia was caused by alcoholic intoxication in three patients and by drug overdose in one patient. Measurements and main results: All patients were rewarmed by a veno-venous bypass and in three cases a convective air warmer was also used. At a bypass flow rate of 100–300 ml/min the mean increase in core temperature was 1.15 °C/h (Range: 1.1–1.2 °C/h). One patient died 2 days after rewarming as a consequence of a reactivated pancreatitis. The other three patients survived without neurological sequelae. Conclusion: This rewarming technique seems safe and effective and allowed the controlled rewarming of our patients who suffered from severe accidental hypothermia
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
  • 10
    ISSN: 1432-1238
    Keywords: Oxygen consumption ; Ventilation, mechanical ; Ventilator weaning ; Post-operative period ; Chronic obstructive pulmonary disease ; Work of breathing
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Notes: Abstract Objective We investigated the effects of continuous positive airway pressure (CPAP) and pressure support ventilation (PSV) on the oxygen cost of breathing ( $$\dot V$$ O2resp) for different states of pulmonary function. Additionally $$\dot V$$ O2resp was measured during spontaneous breathing. Design This was done in a controlled and prospective study. Ventilatory modes were applied randomly. Setting Measurements were performed in a quiet room on volunteers (VOL) and inpatients treated for chronic obstructive pulmonary disease (COPD). Post-operative patients after aortocoronary bypass surgery (ACB) were studied on the cardio-thoracic intensive care unit just before and after extubation. Patients Healthy volunteers (n=14), postoperative patients after aorto-coronary bypass surgery (n=15) and patients with COPD (n=9), xFEV1 47.7%) were the objects of study. Interventions Demand flow CPAP (5 mbar) and PSV (7 mbar, PEEP 5 mbar), using the Hamilton Veolar ventilator, were investigated in comparison to spontaneous breathing. Measurements and results $$\dot V$$ O2 measured by a Datex Deltatrac metabolic monitor. $$\dot V$$ O2resp was calculated by subtraction of total oxygen uptake $$\dot V$$ O2tot) in controlled mode ventilation (CMV) from that in the respective spontaneous breathing mode. For VOL and COPD patients who were not intubated, a CPAP facemask connected to a short 7.5 mm tube was used as connection to the ventilator. Breathing spontaneously under a canopy system VOL showed a VO2resp of 4.5±4.0% compared to 9.2±3.5% for ACB and 15.4±7.7% for COPD. CPAP changed the VO2resp to 7.8±3.9%, 12.0±4.0% and 9.1±3.6% respectively. PSV reduced the $$\dot V$$ O2resp to 7.9±3.8% in ACB and 7.7±5.5% in COPD. Conclusions This investigation confirms findings that postoperative patients have a mild increase in $$\dot V$$ O2resp. COPD exhibit the highest increase in VO2resp. Tracheal tubes, masks and CPAP on a demand flow apparatus increases $$\dot V$$ O2resp in volunteers and postoperative patients after cardiac surgery. The same amount of CPAP in contrary reduces $$\dot V$$ O2resp in patients with COPD. Pressure support ventilation can offset the additional $$\dot V$$ O2resp induced by CPAP but at the same level does not further reduce $$\dot V$$ O2resp in COPD patients.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...