    Springer
    European radiology 6 (1996), S. 838-843 
    ISSN: 1432-1084
    Keywords: Magnetic resonance imaging ; Fast techniques ; Echo-planar imaging ; Focal liver lesions
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Notes: Abstract The performance of breath-hold MR imaging using two T2-weighted hybrid sequences (TSE, TGSE), two T2-weighted single-shot sequences (HASTE, EPI-SE), and one T1-weighted gradientecho sequence (FLASH) was compared with a standard conventional T2-weighted SE sequence in 20 patients with focal liver lesions. Liver signal-to-noise ratio was highest spleen-liver contrast-to-noise ratio (54.3 ± 8.3) and thee HASTE (41.1 ± 12.5) sequence, whereas the highest spleen-liver contrast-noise-ratio was obtained by the TSE sequence (38.9 ± 20.7). Lesion-liver CNR was highest with the TSE sequence (63.9 ± 21.4). With both TSE and HASTE significantly (p 〈 0.01) more lesions were detected as compared with SE and EPE-SE sequences. Our results indicate that breath-hold TSE and HASTE sequences will eventually replace conventional T2-weighted SE techniques due to their insensitivity to motion artifacts, superior lesion detectability and inherently short acquisitions times.